GE Healthcare · CTLightSpeed260007417506901491226000741620053
LightSpeed 5.X (80/100kW)
Symptom
Conflicting warnings appear on the screen with Detector Temperature Is Low then another with Detector Temperature is High. Conflicting warnings appear on the screen with Detector Temperature Is Low then another with Detector Temperature is High..
Error message
Error: 260007417 Detector Temperature Is Low. Image quality may be affected See Long Description for full error Wed Jan 18 07:25:21 2006 Error: 260007417 Host: DAS Process: DCB Firmware File: dtemptst.c Line: 373 Function: DATA ACQUISITION : Detector Heater Scan: 50690/1/2 Type: Axial Exception Level: Sec/Soft Time: 07:25:21:605 Log Series: 51 Detector Temperature Is Low. Image quality may be affected. Current Reading:31.54 Deg C A/D Counts:14912 Error: 260007416 Detector Temperature Is High. Image quality may be affected See Long Description Tue Sep 6 09:07:45 2005 Error: 260007416 Host: orp Process: orp_subsys File: dtemptst.c Line: 364 Tue Sep 6 09:07:44 2005 Host : Das Ermes # : 260007416 Exception Class : Soft Severity : Sec/Soft File : dtemptst.c Line# : 364 Function : No System Function Reported Scan Type : Scout Scan: 372/1/1 Detector Temperature Is High. Image quality may be affected. Current Reading:40.41 Deg C A/D Counts:20053
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2294856-2 (old # 2294856) DHCB assembly Detector Heater Control Board See LD for femc info and picture 2294856-2 (old # 2294856 ) DHCB assembly Comments: FLS - DHCB, BOARD / COVER ASM (H3) DETECTOR HEATER CONTROLLER BD picture Part #: 2294856-2 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: DHCB, Board-Cover Asm EMC ED2 FRU: Yes Status: Orderable Parts Offered as: Current: 2294856-2 Harvested: Not Available Exchange: Not Available Comments: Added per EMC2 RSPL, this part is EMC2 compliant replacement for 2294856. ms13452 050509
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.