LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2

Symptom

Scanner Hardware Stopped Scan HEMRC Errors See Long Description for more information. Scanner Hardware Stopped Scan message appears during Helical Scanning with the following Errors: 185553 (The HEMRC drive has detected a Line Loss - Alarm status: 20H. This occurs when the input power to the drive falls below 85% of the nominal Bus voltage) Possible causes: X-Ray tube stator , HEM-IT, HEMRC power supply 183318 (The HEMRC output current is out of 45.0 percent WARNING tolerance), Monitored state = 2 (2=Run state) Expected current = 2.748 A Actual current = 4.097 A. 183319 (The HEMRC output current is out of 55.0 percent MAXIMUM tolerance), Monitored state = 2 (2=Run state) Expected current = 2.748 A Actual current = 4.653 A. 183316 (The HEMRC failed to reach the Run speed in the allotted time), and 185548 (The HEMRC operating frequency is below the minimum value) Actual frequency: 106.973 Minimum frequency limit: 130.000 At Frequency signal: Inactive Expected: Inactive Rotor State = Transition

Error message

Error: 185553 - Alarm status: 20H (The HEMRC drive has detected a Line Loss. See Long Description for the complete error Tue Aug 10 16:44:06 2004 Error: 185553 Host: OBC Process: Generator Function: XRAY GENERATION : Tube Rotor Control Scan: 34865/2/1 Type: Helical Exception Level: Pri/Most Severe The HEMRC drive has detected a Line Loss. This occurs when the input power to the drive falls below 85% of the nominal Bus voltage. Possible causes: X-Ray tube stator, HEM-IT, HEMRC power supply, .. Alarm status: 20H Or Fri Mar 30 15:21:20 2007 Error: 185553 Function: XRAY GENERATION : Tube Rotor Control Scan: 21538/1/2 Type: Scout Exception Level: Pri/Most Severe The HEMRC drive has detected a Line Loss. This occurs when the input power to the drive falls below 85% of the nominal Bus voltage. Possible causes: - X-Ray tube stator, - HEM-IT, - HEMRC power supply, ... Alarm status: 26H . Note: Alarm Status 26H instead of 20H. This error concept should have additional information for the meaning of the various Hex codes. 20 to 21: Undefined Fault This fault code is undefined and may indicate a defective Drive. Retry operation. If problems persist contact the factory for instructions. or 20: 

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2179860-2 HEMRC CONTROL BOARD See LD for more info 2179860-2 HEMRC CONTROL BOARD. Part #: 2179860-2 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: HEMRC CONTROL BOARD FRU: Yes Status: Orderable FEMC PART COMMENTS: FLS - HEMRC CONTROL BOARD. REPLACES P/N 2179860. ADDED PER PCN 220717.
Area
CT

Full cause & step-by-step resolution