GE Healthcare · CTHiSpeed184823
Scanner pausing during scout
Symptom
Scanner pausing in mid series and not completing the scan prescription
Error message
Error: 184823 Pausing scan because drive position is not known Long Description contains complete error Tues Mar 7 12:44:01 2006 Host: ETC Proc: Cradle Error: 184823 File: cradle_process_rx.c 1.26 Line : 392 Function: No System function Reported Scan 4/4/1 Type: Scout Exception Level : Pri/Most Servere Time: 12:44:01 Log series: 486 Pausing scan because drive position is not known
Applies to
- Models
- HiSpeed
- Product
- HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2263663 Table Can bus I/F cable 2263663 Table Can bus I/F cable Part #: 2263663 Part Type: GP CONSUMABLES Part Weight: 1.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: CAN,COVERS TO ETC FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY: NOT GPO SUPPORTED NOT A FRU. MS26561 DSM 2-12-01 CHANGING FROM FRU=N TO FRU=1 PER PCN 220366. MSA8429 01/30/02
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required