Symptom

Lost control of trackball on operator console during image viewing. OS Kernel aborts and crashes Linux. See Long Description for more Details Lost control of trackball on operator console during image viewing with Error: kernel: keyboard.c: can't emulate rawmode for keycode 272, 273, or 274, then OS Kernel aborts and crashes Linux. [insite@ct01 config]$ tail -420f /var/log/messages Nov 22 13:07:31 ct01 kernel: keyboard.c: can't emulate rawmode for keycode 273 Nov 22 13:07:36 ct01 last message repeated 59 times Nov 22 13:07:37 ct01 kernel: keyboard.c: can't emulate rawmode for keycode 272 Nov 22 13:07:37 ct01 last message repeated 3 times Nov 22 13:07:37 ct01 kernel: keyboard.c: can't emulate rawmode for keycode 274 Nov 22 13:07:41 ct01 last message repeated 7 times [insite@ct01 log]$ egrep "can't|data|error|fail|hard|panic|retry|soft" /var/log/messages* | more /var/log/messages .1:Nov 22 13:55:57 ct01 kernel: keyboard.c: can't emulate rawmode for keycode 273 /var/log/messages.1:Nov 22 13:56:08 ct01 kernel: keyboard.c: can't emulate rawmode for keycode 272 Other Issues seen when this error is found in message log: System takes a long time booting up and is sluggish after the boot up.

Error message

kernel: keyboard.c: can't emulate rawmode for keycode 272 insite@ct01 config]$ tail -420f /var/log/messages Nov 22 13:07:31 ct01 kernel: keyboard.c: can't emulate rawmode for keycode 272

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2376564 USB adapter --old-- used with HP computer. Adapter is NOT needed with new USB Trackball. USB trackball ONLY is 2399753 but if you presently have the adapter you must order the USB trackball UPGRADE KIT 5399753 See LD for more info and pictures old 2376564 USB adapter For trackball used with HP computer, you can now order a USB trackball that does not need this adapter. Part number for USB trackball ONLY is 2399753 but if you presently have the adapter you must order the USB trackball UPGRADE KIT is 5399753 Below picture shows HP using adapter. Below picture shows HP WITHOUT adapter Part #: 5399753 See info below in BLUE for Trackball ONLY Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Kit -USB Trackball and bulkhead with USB cable FRU: Yes Status: Orderable Comments: Added per EMC2 RSPL, ms13452 041201 ms13452 Below is trackball ONLY Part #: 2399753 Part Type: GP CONSUMABLES Part Weight: -67999.32 LBR Repairable: No Returnable: No Hazardous Material: N Description: TRACKBALL - 3 BTTN USB for GRE FRU: Yes Status: Orderable Comments: Added per VCT RSPL, ms13452 041201 Included in FMI 25359, ms13452
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.