Pular para o conteúdo principal
GE Healthcare · MR2248599

Getting Head T/R Faults in Body and Head mode only when the Head coil was plugged into one side of the MSHS.

Symptom

Getting Head T/R Faults in Body and Head mode only when the Head coil was plugged into one side of the MSHS. Getting Head T/R Fault 2249149 "Head path open" intermittently in Body scanning and in Head scanning, but only when the Head coil was plugged into one side of the MSHS. The system is monitoring the Head T/R circuit during Body scanning so it's not surprising that a Head T/R fault could occur when scanning in Body mode.

Error message

2248599 A T/R or dynamic disable fault occurred. 2248599 IC_OPEN_SG NSP APS:aps_cf 220 A T/R or dynamic disable fault occurred.

Applies to

Product
Open Speed with EXCITE For HFO4 and HFO5 only
Software
OPENMR5-HFO5 Open Speed with Excite Product
Component
2254635 RF TRANSMITTER SWITCH This is the (head RF transmit) relay mounted on the back of the GRFD CABINET to switch the head RF transmit line between the left and right sides of the MSHS depending on which side of the MSHS the head coil is plugged into.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues