Pular para o conteúdo principal
GE Healthcare · MRSigna22238172224790222570422253342225337

CWA ID35110 No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet

Symptom

No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.

Error message

2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.

2224790 Hardware Error: Please reset TPS. A MDS link error during download 2224790 Hardware Error: Please reset TPS. A MDS link error during download. The SPI failed to receive a MDS transmission echo. Check the connections and signal level on the MDS fiber optic link. Processor: GRAD File: mdsboot.i96 MDS timer: 0x0000 Xmtd byt

2225704 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Error: 2225704 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_mds Line: 300 MDS link failure. The MDS fiber optic link appears to be broken between the RF and System cabinets. Check RF cabinet power, fiber optic connections and signal level of the fiber optic link between the cabinets. Run MDS fiber optic link diagnostics.

2225334 SPI-SRI Communication Error. SRI failed to respond to a data packet from SPI. Error: 2225334 Host: MRS2 OC0 Process: NSP File: HPC PORT ID:spi sri resp timeout Line: 12 SPI-SRI Communication Error. SRI failed to respond to a data packet from

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required