GE Healthcare · CTHiSpeed18310918763

HiSpeed CT/i 4.X with Indigo and HSA Gantry

Symptom

Scanner Hardware Fault with high voltage spits. Scanner Hardware Fault with high voltage spits

Error message

Error: 183109 KV fault detected: Anode overcurrent See Long Description for complete error Tue Nov 30 15:38:17 2004 Error: 183109 Suite: CT01 Host: OBC Process: Generator File: gen_abort.c 1.33 Method: No Method Line: 248 Function: XRAY GENERATION : KV Loop Scan: 18763/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 15:38:16:120 Log Series: 991 KV fault detected: Anode overcurrent. Fault latch address = ffcff3 hex, Bit = D4

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
45561210 Anode tank used with MX165 Tube on HSA and CTi ---NOT FOR PERFORMIX TUBE--- See LD for compatiblity issues 45561210 Anode tank used with MX165 Tube on HSA and CTi ---NOT FOR PERFORMIX TUBE--- Picture Below 2161305/T3202KB IS BACKWARDS COMPATIBLE AND MAY BE USED TO FILL A DEMAND FOR 45561210/T3202BF IF NO STOCK IS AVAILABLE. A 45561210/ T3202BF CANNOT BE USED TO REPLACE A 2161305/T3202KB Part #: 45561210 Part Type: GP REPAIRABLES Part Weight: 0.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ANODE HV SUPPLY FRU: Yes Status: Orderable Comments: FLS - H: 45561210/T3202BF IS REPAIRED IN FRANCE. IF NO STOCK AVAILABLE, A 2161305/ T3202KB CAN BE SUBSTITUTED. MS4636 5/15/97 DO NOT SUBSTITUTE T3202BF FOR A T3202KB, THE T3202BF IS NOT FORWARD COMPATIBLE
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.