The System Fails To Scan With An SGA PS Overcurrent Fault.

Symptom

Will Not Scan with Gradient Errors.

Error message

2244590 SGA-PS OverCurrent Fault 2244590 "SGA-PS OverCurrent Fault.\n\ Look at SGA-PS OC front LEDs to determine which SGA axis.\n\ If primary OC, replace SGA-PS. If secondary OC, swap military\n\ connectors to isolate SGA vs. SGA-PS. Power cycle ACGD POWER\n\ PDU breaker to recover."

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2270055-2 or T3504MG ACGD SGA Module ACGD SGA Module 2270055-2 or T3504MG. Note: You can see SGA Amplifiers on HDx systems when they were upgraded to HDX from 11.0. The original SGA's were kept and only the Control boards were upgraded. ACGD SGA Module 2270055-2 or T3504MG. Note: You can see SGA Amplifiers on HDx systems when they were upgraded to HDX from 11.0. The original SGA's were kept and only the Control boards were upgraded.