LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Unable to resume scanning right away after pausing. Scan will abort posting errors See Long Description for error code examples. Unable to resume scanning right away after pausing. System aborts scan attempts posting various generator errors including 260118262 (Jedi fault 80-1425) in gesyslog. Problem occurred after operator paused a scan for some reason and then attempted to resume quickly. System posted approx 5 scan aborts in gesyslog upon resume attempt but then finally started again after about a 90 second delay. Several Jedi fault errors were posted in gesylog, with fault 80-1425 repeated several times. Error stream seen in gesylog: Tue Jul 5 01:03:28 2005 Error: 245040 Host: hactgels1 Process: scanmgr File: scanmgr_events.cxx Line: 1629 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 260000 Operator Aborted Scanning Tue Jul 5 01:03:28 2005 Error: 200281001 Host: hactgels1 Process: visScript File: VavCTQueueMgr.cxx Line: 390 Status update: Session 1 cancelled: exam=999, series=2, seqnum=16 Tue Jul 5 01:03:29 2005 Error: 260118236 Host: orp Process: orp_subsys File: gen_exec.c Line: 2161 Tue Jul 5 01:03:28 2005 Host : Jedi Ermes # : 260118236 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2161 Function : No System Function Reported Scan Type : None/Unknown Scan: 999/2/0 (70-0501)DC bus out of range.^C Z6,Z7= 86h,1h Tue Jul 5 01:03:29 2005 Error: 260118167 Host: orp Process: orp_subsys File: gen_exec.c Line: 2161 Tue Jul 5 01:03:29 2005 Host : Jedi Ermes # : 260118167 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2161 Function : No System Function Reported Scan Type : None/Unknown Scan: 999/2/0 (40-0104)Rotation current open load. Rotation board detected that no current is flowing to the motor. ^C Z6,Z7= 7h,0h Tue Jul 5 01:03:56 2005 Error: 200281001 Host: hactgels1 Process: visScript File: VavCTQueueMgr.cxx Line: 220 Status update: Start session 1, context 4: exam=999, series=2, seqnum=17, first img=1, numslices=306 slice-center=(-13.500000,0.000000), slice-range=[314.500000:-66.750000], dfov=408.000000, slice-thickness=2.500000, gantry-tilt=0.000000 Tue Jul 5 01:03:58 2005 Error: 260118262 Host: orp Process: orp_subsys File: gen_exec.c Line: 2161 Tue Jul 5 01:03:58 2005 Host : Jedi Ermes # : 260118262 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2161 Function : No System Function Reported Scan Type : Helical Scan: 999/2/1 (80-1425)Time since last rotor acceleration is insufficent. Probable software error. Z6,Z7= 0h,0h Tue Jul 5 01:03:58 2005 Error: 245043 Host: hactgels1 Process: scanmgr File: scanmgr_events.cxx Line: 1641 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 262000 Scanning System Aborted Scanning Tue Jul 5 01:03:58 2005 Error: 245213 Host: hactgels1 Process: scanmgr File: scanmgr_events.cxx Line: 1647 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 262000 Scan was Paused/Aborted by Firmware Tue Jul 5 01:03:59 2005 Error: 245190 Host: hactgels1 Process: scanmgr File: scan_complete.cxx Line: 336 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 262000 System aborted Helical scanning. No views, either offset or data, were collected/received. Data views expected: 10688 Tue Jul 5 01:03:59 2005 Error: 200281001 Host: hactgels1 Process: visScript File: VavCTQueueMgr.cxx Line: 390 Status update: Session 1 cancelled: exam=999, series=2, seqnum=17 Tue Jul 5 01:04:10 2005 Error: 200281001 Host: hactgels1 Process: visScript File: VavCTQueueMgr.cxx Line: 220 Status update: Start session 1, context 4: exam=999, series=2, seqnum=18, first img=1, numslices=306 slice-center=(-13.500000,0.000000), slice-range=[314.500000:-66.750000], dfov=408.000000, slice-thickness=2.500000, gantry-tilt=0.000000 Tue Jul 5 01:04:10 2005 Error: 260118262 Host: orp Process: orp_subsys File: gen_exec.c Line: 2161 Tue Jul 5 01:04:10 2005 Host : Jedi Ermes # : 260118262 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2161 Function : No System Function Reported Scan Type : Helical Scan: 999/2/1 (80-1425)Time since last rotor acceleration is insufficent. Probable software error.^C Z6,Z7= 0h,0h Tue Jul 5 01:04:10 2005 Error: 245043 Host: hactgels1 Process: scanmgr File: scanmgr_events.cxx Line: 1641 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 264000 Scanning System Aborted Scanning Tue Jul 5 01:04:10 2005 Error: 245213 Host: hactgels1 Process: scanmgr File: scanmgr_events.cxx Line: 1647 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 264000 Scan was Paused/Aborted by Firmware Tue Jul 5 01:04:10 2005 Error: 245190 Host: hactgels1 Process: scanmgr File: scan_complete.cxx Line: 336 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 264000 System aborted Helical scanning. No views, either offset or data, were collected/received. Data views expected: 10688 Tue Jul 5 01:04:11 2005 Error: 200281001 Host: hactgels1 Process: visScript File: VavCTQueueMgr.cxx Line: 390 Status update: Session 1 cancelled: exam=999, series=2, seqnum=18 Tue Jul 5 01:04:16 2005 Error: 200281001 Host: hactgels1 Process: visScript File: VavCTQueueMgr.cxx Line: 220 Status update: Start session 1, context 4: exam=999, series=2, seqnum=19, first img=1, numslices=306 slice-center=(-13.500000,0.000000), slice-range=[314.500000:-66.750000], dfov=408.000000, slice-thickness=2.500000, gantry-tilt=0.000000 Tue Jul 5 01:04:16 2005 Error: 260118262 Host: orp Process: orp_subsys File: gen_exec.c Line: 2161 Tue Jul 5 01:04:16 2005 Host : Jedi Ermes # : 260118262 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2161 Function : No System Function Reported Scan Type : Helical Scan: 999/2/1 (80-1425)Time since last rotor acceleration is insufficent. Probable software error.^C Z6,Z7= 0h,0h Tue Jul 5 01:04:16 2005 Error: 245043 Host: hactgels1 Process: scanmgr File: scanmgr_events.cxx Line: 1641 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 266000 Scanning System Aborted Scanning Tue Jul 5 01:04:16 2005 Error: 245213 Host: hactgels1 Process: scanmgr File: scanmgr_events.cxx Line: 1647 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 266000 Scan was Paused/Aborted by Firmware Tue Jul 5 01:04:16 2005 Error: 245190 Host: hactgels1 Process: scanmgr File: scan_complete.cxx Line: 336 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 266000 System aborted Helical scanning. No views, either offset or data, were collected/received. Data views expected: 10688 Tue Jul 5 01:04:17 2005 Error: 200281001 Host: hactgels1 Process: visScript File: VavCTQueueMgr.cxx Line: 390 Status update: Session 1 cancelled: exam=999, series=2, seqnum=19 Tue Jul 5 01:04:22 2005 Error: 200281001 Host: hactgels1 Process: visScript File: VavCTQueueMgr.cxx Line: 220 Status update: Start session 1, context 4: exam=999, series=2, seqnum=20, first img=1, numslices=306 slice-center=(-13.500000,0.000000), slice-range=[314.500000:-66.750000], dfov=408.000000, slice-thickness=2.500000, gantry-tilt=0.000000 Tue Jul 5 01:04:23 2005 Error: 260118262 Host: orp Process: orp_subsys File: gen_exec.c Line: 2161 Tue Jul 5 01:04:22 2005 Host : Jedi Ermes # : 260118262 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2161 Function : No System Function Reported Scan Type : Helical Scan: 999/2/1 (80-1425)Time since last rotor acceleration is insufficent. Probable software error.^C Z6,Z7= 0h,0h Tue Jul 5 01:04:23 2005 Error: 245043 Host: hactgels1 Process: scanmgr File: scanmgr_events.cxx Line: 1641 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 268000 Scanning System Aborted Scanning Tue Jul 5 01:04:23 2005 Error: 245213 Host: hactgels1 Process: scanmgr File: scanmgr_events.cxx Line: 1647 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 268000 Scan was Paused/Aborted by Firmware Tue Jul 5 01:04:23 2005 Error: 245190 Host: hactgels1 Process: scanmgr File: scan_complete.cxx Line: 336 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 268000 System aborted Helical scanning. No views, either offset or data, were collected/received. Data views expected: 10688 Tue Jul 5 01:04:23 2005 Error: 200281001 Host: hactgels1 Process: visScript File: VavCTQueueMgr.cxx Line: 390 Status update: Session 1 cancelled: exam=999, series=2, seqnum=20 Tue Jul 5 01:04:27 2005 Error: 200281001 Host: hactgels1 Process: visScript File: VavCTQueueMgr.cxx Line: 220 Status update: Start session 1, context 4: exam=999, series=2, seqnum=21, first img=1, numslices=306 slice-center=(-13.500000,0.000000), slice-range=[314.500000:-66.750000], dfov=408.000000, slice-thickness=2.500000, gantry-tilt=0.000000 Tue Jul 5 01:04:27 2005 Error: 260118262 Host: orp Process: orp_subsys File: gen_exec.c Line: 2161 Tue Jul 5 01:04:26 2005 Host : Jedi Ermes # : 260118262 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2161 Function : No System Function Reported Scan Type : Helical Scan: 999/2/1 (80-1425)Time since last rotor acceleration is insufficent. Probable software error.^C Z6,Z7= 0h,0h Tue Jul 5 01:04:27 2005 Error: 245043 Host: hactgels1 Process: scanmgr File: scanmgr_events.cxx Line: 1641 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 270000 Scanning System Aborted Scanning Tue Jul 5 01:04:27 2005 Error: 245213 Host: hactgels1 Process: scanmgr File: scanmgr_events.cxx Line: 1647 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 270000 Scan was Paused/Aborted by Firmware Tue Jul 5 01:04:27 2005 Error: 245190 Host: hactgels1 Process: scanmgr File: scan_complete.cxx Line: 336 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 270000 System aborted Helical scanning. No views, either offset or data, were collected/received. Data views expected: 10688 Tue Jul 5 01:04:28 2005 Error: 200281001 Host: hactgels1 Process: visScript File: VavCTQueueMgr.cxx Line: 390 Status update: Session 1 cancelled: exam=999, series=2, seqnum=21 Tue Jul 5 01:04:31 2005 Error: 200281001 Host: hactgels1 Process: visScript File: VavCTQueueMgr.cxx Line: 220 Status update: Start session 1, context 4: exam=999, series=2, seqnum=22, first img=1, numslices=306 slice-center=(-13.500000,0.000000), slice-range=[314.500000:-66.750000], dfov=408.000000, slice-thickness=2.500000, gantry-tilt=0.000000 Tue Jul 5 01:04:31 2005 Error: 260118262 Host: orp Process: orp_subsys File: gen_exec.c Line: 2161 Tue Jul 5 01:04:31 2005 Host : Jedi Ermes # : 260118262 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2161 Function : No System Function Reported Scan Type : Helical Scan: 999/2/1 (80-1425)Time since last rotor acceleration is insufficent. Probable software error.^C Z6,Z7= 0h,0h Tue Jul 5 01:04:31 2005 Error: 245043 Host: hactgels1 Process: scanmgr File: scanmgr_events.cxx Line: 1641 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 272000 Scanning System Aborted Scanning Tue Jul 5 01:04:31 2005 Error: 245213 Host: hactgels1 Process: scanmgr File: scanmgr_events.cxx Line: 1647 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 272000 Scan was Paused/Aborted by Firmware Tue Jul 5 01:04:32 2005 Error: 245190 Host: hactgels1 Process: scanmgr File: scan_complete.cxx Line: 336 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 272000 System aborted Helical scanning. No views, either offset or data, were collected/received. Data views expected: 10688 Tue Jul 5 01:04:32 2005 Error: 200281001 Host: hactgels1 Process: visScript File: VavCTQueueMgr.cxx Line: 390 Status update: Session 1 cancelled: exam=999, series=2, seqnum=22 Tue Jul 5 01:05:01 2005 Error: 200281001 Host: hactgels1 Process: visScript File: VavCTQueueMgr.cxx Line: 220 Status update: Start session 1, context 4: exam=999, series=2, seqnum=23, first img=1, numslices=306 slice-center=(-13.500000,0.000000), slice-range=[314.500000:-66.750000], dfov=408.000000, slice-thickness=2.500000, gantry-tilt=0.000000 Tue Jul 5 01:05:02 2005 Error: 260118180 Host: orp Process: orp_subsys File: gen_exec.c Line: 2161 Tue Jul 5 01:05:01 2005 Host : Jedi Ermes # : 260118180 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2161 Function : No System Function Reported Scan Type : Helical Scan: 999/2/1 (40-0162)Acceleration while in error. ^C Z6,Z7= 0h,0h Tue Jul 5 01:05:02 2005 Error: 245043 Host: hactgels1 Process: scanmgr File: scanmgr_events.cxx Line: 1641 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 274000 Scanning System Aborted Scanning Tue Jul 5 01:05:02 2005 Error: 245213 Host: hactgels1 Process: scanmgr File: scanmgr_events.cxx Line: 1647 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 274000 Scan was Paused/Aborted by Firmware Tue Jul 5 01:05:02 2005 Error: 245190 Host: hactgels1 Process: scanmgr File: scan_complete.cxx Line: 336 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 274000 System aborted Helical scanning. No views, either offset or data, were collected/received. Data views expected: 10688 The following error was seen in the error stream on another system where this issue occurred: Wed Apr 20 14:45:38 2005 Error: 260118324 Host: orp Process: orp_subsys File: gen_exec.c Line: 3937 Wed Apr 20 14:45:37 2005 Host : On-Board Rotating Processor Ermes # : 260118324 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 3937 Function : XRAY GENERATION : Tube Rotor Control Scan Type : Helical Scan: 90/2/1 Unable to bring up rotor due to acceleration interval constraint. Remaining Time: 83 seconds RotState:0 Mode:0x3 RmExpD:15000 Accel:6960 (Rx:20 Last brake:6585 Decel:5950 EndExam:1799740 NoRxes:5950 Tmrs: Rotor Idle:0 Restore:0 Fil Idle:0 Restore:0)
Error message
Error: 260118262 (80-1425)Time since last rotor acceleration is insufficent. Probable software error. Tue Jul 5 01:03:58 2005 Error: 260118262 Host: orp Process: orp_subsys File: gen_exec.c Line: 2161 Tue Jul 5 01:03:58 2005 Host : Jedi Ermes # : 260118262 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2161 Function : No System Function Reported Scan Type : Helical Scan: 999/2/1 (80-1425)Time since last rotor acceleration is insufficent. Probable software error. Tue Jul 5 01:03:58 2005 Error: 260118262 Host: orp Process: orp_subsys File: gen_exec.c Line: 2161 Tue Jul 5 01:03:58 2005 Host : Jedi Ermes # : 260118262 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2161 Function : No System Function Reported Scan Type : Helical Scan: 999/2/1 (80-1425)Time since last rotor acceleration is insufficent. Probable software error. Z6,Z7= 0h,0h
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Software
- 05MW10.2 for LightSpeed 7.X with Xtream FX GOC5 Console - check_config Example: 05MW10.2.V40_H_V64_G_GTL VCT Software 05MW10.2 M3 Release P/N Description 5135205 M3 Operating System 5135848 M3 H-64/32 Apps 1 5135849 M3 H-64/32 Apps 2 5135850 M3 DARC Apps 5135851 M3 Class C 5135852 M3 Class M 05MW10.2 for LightSpeed VCT 7.X with Xtream FX GRE-GOC5 Console Introduced with LS VCT M3 systems produced from March, 2005 Part Number Description Version Showprods 5135205 M3 Operating System CTT 4.3.10 05MW10.2 5135848 M3 H-64/32 Apps 1 vctme2.1E 05MW10.2 5135849 M3 H-64/32 Apps 2 vctme2.1E 05MW10.2 5135850 M3 DARC Apps vctme2.1E 05MW10.2 5135851 M3 Class C vctme2.1E_c 05MW10C.2 5135852 M3 Class M vctme2.1E_m 05MW10M.2 Options: B7864KE CardIQ Pro for Operator Console (Pre-requisite B7540LD) B7864KF CardIQ Plus for Operator Console (Pre-requisite B7540LD) B7864KL CTC Pro for Operator Console (Pre-requisite B7540LD) B7868EP CardEP for Operator Console (Pre-requisite B7540LD) B7700PY CT Perfusion 3-Neuro Package for Operator Console B7700PZ Perfusion 3-Multi Organ for CT Operator Console B7868AB AutoBone for CT Operator Console (Pre-requisite B7540LD) B7540LS DentaScan Option for the LightSpeed Operator’s Console B7710LS CardIQ SnapShot Option for the LightSpeed Multi-slice scanners B7540LD Volume Viewer for Operator Console (VA2, VA2, NAV2) B7868LA Advanced Vessel Analysis for Operator Console (Pre-requisite B7540LD) check_config Example: 05MW10.2.V40_H_V64_G_GTL 05MW10.2.V40_H_V64_G_GTL - equates to the M3 software release currently in production as of April 2005 How do you decode this: 05 M W10.2. V40 H V64 G GTL - 20 05 - M ilwaukee - W eek FW-10 - 2 number of Respins - V40 Detector width = 40 mm - H Hecules tube - V64 Das Type 64 Slice - G = GRE - GTL Global Table (2000) or GTM (1700) The Engineering team talks in build levels ! Question: How to determine the Build level of the VCT software?? [insite@ct04 ~]$ what /usr/g/bin/scanRx helios: Copyright @ 2005 GE Medical Systems helios: Build: 20050218:0020 helios: Component: scanRx helios: Version: 20050218:0020 helios: Date of Build: Fri Feb 18 00:20:38 CST 2005 helios: System: Linux liberty15.ct.med.ge.com 2.4.20-28.9.XFS1.3.1s mp #1 SMP Mon Jan 5 13:20:15 CST 2004 i686 i686 i386 GNU/Linux helios: Working directory: /vobs/uif/ExamRx/ScanRx/src helios: Search path: /data/stage/nightly/liberty/vct.94.20050217213 000Thu/BuildScripts:/vobs/helios/com_util/bin/Linux2.4:/vobs/helios/util/scr ipts:/usr/atria/bin:/sbin:/usr/bin:/bin:/usr/bin/X11:/usr/lib/Zmail/bin:/usr /etc:/usr/bsd:/usr/3p/gebin:/usr/openwin/bin:/opt/SUNWspro/bin:/usr/ccs/bin: /usr/ucb:/usr/java/bin:.:/usr/local/bin:/usr/3p/vbroker/vbc/bin:/usr/3p/vbro ker/vbj/bin helios: Original View (not winked-in): vct.94 helios: element /vobs/gsp/... TERRA_CT_20050215 -nocheckout helios: element /vobs/terra/... TERRA_CT_20050215 -nocheckout helios: element /vobs/gvtk/... GVTK_20050113 -nocheckout helios: Builder: ctbuild helios: Config Spec: helios: element * VCT.94 DbNumbers.c1.1 4/21/94 16:14:03 Copyr 1991 GE {ctuser@ct04}[5] cat ~service/state/CurrSysCfgInfo Hospital_Name=FROEDTERT HOSP. MILW. WI USN=0000353967 MLN=9999 System_ID=414805CT4 Product_Line=LightSpeed VCT Modality=CT Suite_Name=CT04 Application_Software_Install_Date=Fri Mar 11 19:59:50 2005 Application_Software_Release_ID=05MW10.2 Base_IOS_Software_Release_ID=IOS_20050113_R293a_P2 IIP_Software_Release_ID=3.3.4-Linux9 Gantry_Type=vct_40 Recon_Hardware_Type=GRE DAS_Type=VDAS64 [root@ct04]# check_config Product: LightSpeed VCT sw_version: sw_hw_string: 05MW10.2.V40_H_V64_G_GTL machine_type: Intel host_id: 39754912 tube_type: 16 gantry_type: GANTRY_VCT_40 No. of processors: 4 Processor speed: 2658 memory: 4096 [insite@ct04 ~]$ showprods | grep -i product I Helios_H2_O2LESS_i18n_OC_Product Helios ver 05MW10 rel 05MW10.2 I HPower_CLASSC_Product HPower ver 05MW10C rel 05MW10C.2 I Helios_OC_Product LightSpeed ver 05MW10 rel 05MW10.2 I Helios_CLASSC_Product CLASS ver 05MW10C rel 05MW10C.2 I Helios_i18n_Product Helios ver 05MW10 rel 05MW10.2 I HPower_OCROOT_Product HPower ver 05MW10 rel 05MW10.2 I Helios_H2_O2LESS_CLASSC_Product Helios ver 05MW10C rel 05MW10C.2 I Helios_CLASSMOC_Product CLASSMROOT ver 05MW10M rel 05MW10M.2 I Helios_H2_O2LESS_OC_Product Helios ver 05MW10 rel 05MW10.2 I Helios_VCT_Product VCT ver 05MW10 rel 05MW10.2 I Helios_H2_PEGASUS_CLASSC_Product Helios ver 05MW10C rel 05MW10C.2 I Helios_GRE_OC_Product GRE ver 05MW10 rel 05MW10.2 I Helios_VCT_CLASSC_Product VCT ver 05MW10C rel 05MW10C.2
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.