GE Healthcare · CTLightSpeed26000738650056260007422
LightSpeed 4.X - GRE
Symptom
Scanner Hardware Stopped Scan with DAS parity errors
Error message
Error: 260007386 DAS Has Detected An Error. Data Parity Error detected in the xx Data Stream(s) on Converter Board xx or the DCB Board. See LD for sample errors Wed Mar 23 15:46:11 2005 Error: 260007386 Host: DAS Process: DCB Firmware File: parity.c Line: 219 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 214/5/1 Type: Helical Exception Level: Sec/Soft Time: 15:46:11:545 Log Series: 660 SCAN ABORT -- DAS Has Detected An Error. Data Parity Error detected in the 1B Data Stream (s) on Converter Board 33 or on Converter Board 41 . NOTE: Before troubleshooting, go through gesys.log (viewlog) and search for all “Data Parity Error” reports in the log. There are 2 items you need to note from the error: 1- Data Parity Error detected in the xx Data Stream and 2- On Converter Board xx or on Converter Board xx Tue Jan 10 09:37:14 2006 Error: 260007386 Host: orp Process: orp_subsys File: parity_16.c Line: 255 Exception Class : Abort Severity : Secondary Function : No System Function Reported Scan Type : Static Scan: 50056/1/1 SCAN ABORT -- DAS Has Detected An Error. Data Parity Error detected in the 3A Data Stream (s) on Converter Board 55 or on Converter Board 71 . Note: Error 260007
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2355552 MDAS 16 Convertor Card ( replaces 2322994-2 ) Used In LightSpeed 16 Gantry with MDAS. 2355552 MDAS 16 Converter Card used in LS 4X with 16 slice MDAS (replaces 2322994-2) Part #: 2355552 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.25 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: SERVICE, MDAS16 CONV BD FRU: Yes Status: Orderable Comments: Please consider ordering the used part when it is appropriate for your customer. It will save you substantial money. This is an MDAS16 CONV BD .compared to the SDAS. Make sure the correct Convertor is Ordered and replaced.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.