Pular para o conteúdo principal
GE Healthcare · MRSigna22367232233396

A power surge at the site took out the RECEIVER board in the ISE chassis

Symptom

Unable to Scan DWI

Error message

2236723 CERD Diags: Fast Receiver Analog Loopback Test Failure. 2236723 nsi-mr diags diagsMain.cc 2332 CERD Diags: Fast Receiver Analog Loopback Test Failure. The test timed out after 5 sec waiting for XD bus data. Check Fast Receiver power and connections. Exp: 512 XD pts. Rcvd: 0 XD pts. FRU(s): CERD Dig Bd, Fast Rcvr Bds + Cables

2233396 The Frame Data Buffer Queue has overflowed. May need a longer TR time, or more time between echoes. Please restart scan. Thu Feb 22 17:59:07 2007 Error: 2233396 Host: MENEMR Process: NSP File: TPS:cap: c_int45 Line: 4 The Frame Data Buffer Queue has overflowed. May need a longer TR time, or more time between echoes. Please restart scan.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2148437 RECEIVER board The part description is "RECEIVER FOR LIGHTNING PROGRAM". This is the board in the ISE chassis which interfaces the Fast Receiver module.
Area
MR