Auto Prescan failed, RF amplifier is not ready.

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2224666 Auto Prescan Failed, RF Amplifier is Not Ready NOTE: When using this error to create new knowledge, please include any associated error codes. 2224666 is a generic error code which requires further information. Mon May 17 08:19:08 2004 Error: 2224666 Host: lx_mr Process: psc File: PscMainControl.cc Line: 1296 Auto Prescan failed, RF amplifier is not ready.

2225656 ERBTEC RF Amplifier fault 77 PA Grid Current Too High Mon May 17 08:18:25 2004 Error: 2225656 Host: lx_mr Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3060 ERBTEC RF Amplifier fault 77. PA grid current too high. Possible defective PA tube, or problems on the power control board or processor board.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR