GE Healthcare · MRSigna2225656
clarify 77 faults Due to PA Tube.
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2225656 ERBTEC RF Amplifier fault 77 PA Grid Current Too High Mon May 17 08:18:25 2004 Error: 2225656 Host: lx_mr Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 3060 ERBTEC RF Amplifier fault 77. PA grid current too high. Possible defective PA tube, or problems on the power control board or processor board.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.