GE Healthcare · CTLightSpeedVCT260100096
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Intermittent Scan Hardware fault
Error message
Error: 260100096 Exposure Interlock Opened See Long Description for complete error. 260100096 Exposure Interlock Opened Mon Dec 6 09:54:17 2004 Error: 260100096 Host: tgp Process: tgp_subsys File: scTrigSeqActions.c Line: 546 Mon Dec 6 09:54:17 2004 Host : Table Gantry Processor Ermes # : 260100096 Exception Class : Abort Severity : Primary File : scTrigSeqActions.c Line# : 546 Function : System Monitoring : Xray Exposure Interlocks Scan Type : Axial Scan: 7629/2/1 Exposure Interlock Opened. The current loop from Gantry to PDU. The cause of the failure: Console Relay
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 2367805 bulkhead includes the FIBER OPTIC cable GRE bulkhead to Darc/dip card, 915 mm / 3 ft. 2367805 bulkhead includes the FIBER OPTIC cable GRE bulkhead to Darc/dip card, 915 mm / 3 ft. See LD for picture 2367805 bulkhead includes the FIBER OPTIC cable GRE bulkhead to Darc/dip card, 915 mm / 3 f t. Part #: 2367805 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: OUTLET BOX ASSY - GOC/GRE FRU: Yes Status: Orderable Comments: FLS - CABLE BULKHEAD ASSY (GRE) USED ON H16 ADDED PER GRE RSPL MS13452 RISK: FUNCITONALITY FREQUENCY: LOW Mark Number FRU Description 4 2367805 Yes CABLE BULKHEAD ASSY (GRE)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.