GE Healthcare · MR2240243
Will Not Scan with Gradient Errors
Symptom
Will Not Scan with Gradient Errors.
Error message
2240243 X Gradient Amplifier Fault. Gradient amplifier fault, over current or clock loss internal to Gradient Subsystem See Long Description Fri Jan 28 07:30:43 2005 Error: 2240243 Host: lsdi Process: NSPFile: HPC_PORT_ID:spi_gca_error Line: 9 X Gradient Amplifier Fault. Gradient amplifier fault, over current or clock loss internal to Gradient Subsystem (GS). Disconnect X amp bus cable and power on GS in local mode. If GS turns on, replace X Gradient Amp;else, check amp bus cable or SCA.
Applies to
- Product
- Profile 3 Ultrasparc CPU with AN8290Y gradients
- Software
- 7.7x Profile3 through 4, Contour3 through 4
- Area
- MR
Full cause & step-by-step resolution