GE Healthcare · MR2240243

Will Not Scan with Gradient Errors

Symptom

Will Not Scan with Gradient Errors.

Error message

2240243 X Gradient Amplifier Fault. Gradient amplifier fault, over current or clock loss internal to Gradient Subsystem See Long Description Fri Jan 28 07:30:43 2005 Error: 2240243 Host: lsdi Process: NSPFile: HPC_PORT_ID:spi_gca_error Line: 9 X Gradient Amplifier Fault. Gradient amplifier fault, over current or clock loss internal to Gradient Subsystem (GS). Disconnect X amp bus cable and power on GS in local mode. If GS turns on, replace X Gradient Amp;else, check amp bus cable or SCA.

Applies to

Product
Profile 3 Ultrasparc CPU with AN8290Y gradients
Software
7.7x Profile3 through 4, Contour3 through 4
Area
MR