GE Healthcare · CTLightSpeed23000815514681230011067
LightSpeed 5.X (60kW)
Symptom
Scanner performs scout scan but no axial or helical. Scanner performs scout scan but no axial or helical .
Error message
Error: 230008155. During start-up processing, a scan header file was found on the system disk without a corresponding view data file on the high-speed disk. See LD for complete error Error: 230008155 Function: System Control : Scan Database Manager Scan Type: Not scanning Scan: 14681/2/1 During start-up processing, a scan header file was found on the system disk without a corresponding view data file on the high-speed disk. Scan: AJAX/14681/2/2 Error: 230011067 Unexpected file, or directory, detected in scan data root directory on scan data disks during start-up. The unexpected file, or directory,will be removed. See LD for complete error Error: 230011067 Unexpected file, or directory, detected in scan data root directory on scan data disks during start-up. The unexpected file, or directory,will be removed. Fri Feb 10 05:49:23 2006 Error: 230011067 Host: ct01 Process: scanfilemgr File: sdio xfs api.c Line: 3500 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Unexpected file, or directory, detected in scan data root directory on scan data disks during start-up. The unexpected file, or directory, will be removed. File/directory name: lost+found Root directo
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 05MW14.5 for all systems with GRE/Xtream consoles on HiSpeed QX/i and LightSpeed 1.X, 2.X, 3.X, 4.X, 5.X Series(QXi, PLUS, ULTRA, LS16, PRO 16, RT). Obsoleted by FMI 25365 deploy FMI 25365 It is the Service person's responsibility to use the appropriate documentation when performing and verifying service. Refer to FMI 25365: 06MW03.5 SW Update for CT-LS 1X/2X/3X/4 DIR 5161945-100, which can be found on the Distributed Technical CDROM/DVD, or online Links to any needed documentation is found in the Product concepts as an attachment called "ProductDocumentLinks"
- Area
- CT