CWA ID3245. Cant Scan A Shoulder, Gradient X Y Z Faults.

Symptom

Will Not Scan with Gradient Errors.

Error message

2244597 The Gradient System on the Z axis Recovered Successfully X, Y or Z axis Faults Fri May 7 17:11:00 2004 Error: 2244597 Host: NSM6 Process: NSP File: SCP:GP Line: 1069 The gradient system on the Z axis recovered successfully. There are no faults at the present time.

2244591 SGA Power Supply Over Voltage Fault See Long Description. Fri May 7 16:14:56 2004 Error: 2244591 Host: NSM6 Process: NSP File: SCP:GP Line: 1857 SGA Power Supply Over Voltage Fault. FOR FIELD ENGINEER: Look at SGA-PS front LEDs. If BUS OV LED lit, check PDU 420V output. If OV X,Y,Z LED lit, swap military connectors to isolate SGA vs. SGA-PS. Power cycle ACGD POWER breaker on PDU to recover.

Applies to

Models
Excite, Discovery, Optima
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.