GE Healthcare · CTLightSpeed26011631950132
LightSpeed 4.X - GRE
Symptom
When pressing Alignment Light Button, Gantry rotates but no lasers come on and the ORP reset indicator flashes on the gantry display
Error message
Error: 260116319 State Response Timeout Occurred See Long Description for full error Fri Sep 2 07:22:58 2005 Error: 260116319 Host: tgp Process: tgp_subsys File: subsys_mgr.c Line: 2343 Fri Sep 2 07:22:55 2005 Host : Table Gantry Processor Ermes # : 260116319 Exception Class : Secondary Severity : Pri/Soft File : subsys_mgr.c Line# : 2343 Function : No System Function Reported Scan Type : None/Unknown Scan: 50132/1/0 State Response Timeout Occurred Subsystem: Table Gantry Processor State: 12 Non-Responding App Id: 100 Timeout Value: 1000 msec.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2105617-5 (replace 2105617-4) SINGLE, TERMINATED LASER DIODE ASM - used on H2 and HP gantry See long description for details SINGLE, TERMINATED LASER DIODE ASSEMBLY Part #: 2105617-5 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: SINGLE, TERMINATED LASER DIODE AMP CONNECTOR, 105MM LONG, PHOTONIC FRU: Yes Status: Orderable Part #: 2105617-4 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: SINGLE, TERMINATED LASER DIODE AMP CONNECTOR, 105MM LONG FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY. COMMON TO MFG MS26561 DSM 1/7/01 PART OF 2229971
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.