Pular para o conteúdo principal
GE Healthcare · MRSignaOvationExcite222569022246662225693

Neither Auto nor Manual Prescan Will Start, No Knocking

Symptom

Auto and Manual Prescan Fails

Error message

2225690 MDS link failure. Error: 2225690 Host: lx-gs Process: NSP File: HPC_PORT_ID:spi_mds Line: 500 MDS link failure. A receive error was detected on the echo of an MDS transmission. Check the connections and signal level on the MDS fiber optic link.

2224666 Auto Prescan failed, SPI-MDS link is not working.

2225693 MDS link failure. An MDS device failed to respond to IPG. Wed Oct 29 00:14:08 1997 2225693 1 CMR1 OC0 HPC_PORT_ID PMH.m 1.29 spi_mds 503 MDS link failure. An MDS device failed to respond to IPG. If the link is broken, a separate error message will be logged. If not, a spurious error occurred. MDS timer: 0x0252 Transmitted header: 0x10 Transmitted opcode: 0x02

Applies to

Models
Signa, Ovation, Excite
Product
Signa Horizon (5X Horizon) 5X Horizon systems
Area
MR