Pular para o conteúdo principal
GE Healthcare · MRSigna2254215

32130 Will Not Scan Due to RF Amplifier Gate Fault 2254215

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2254215 SRFD2 RF Amplifier Gate Fault. Error: 2254215 SRFD2 RF Amplifier Gate Fault. RF Amplifier reports an unblank pulse width exceeding specification (100sec). Possible causes: the PSD is exceeding the limits of the system, or defective RF Amplifier.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required