Pular para o conteúdo principal
GE Healthcare · CTHiSpeed178574178568

Scanning hardware reset fails

Symptom

Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .

Error message

Error: 178574 The following controllers STC/ETC/OBC failed to report initialization status in 120 seconds See Long description for error code variations Sun Nov 27 14:18:24 2005 Error: 178568 Suite: SBCT Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to ETC in 10 seconds ------------------------------------------------------------- Wed Sep 8 06:20:02 2004 Error: 178568 Host: cmls Process: fwmgr File: NBSSocket.c Line: 388 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to establish socket connection to OBCR in 10 seconds. ---------------------------------------------------------------------------- Fri Dec 24 04:38:04 2004 Error: 178568 Host: ct02_sbc Process: fwmgr File: NBSSocket.c Line: 388 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to establish socket connection to STC in 10 seconds

Applies to

Models
HiSpeed
Product
HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.