GE Healthcare · MRExcite224821728471742248163
Receiver Cal Errors Logged on Every System Boot
Symptom
Receiver Cal Errors logged on every System Boot Errors logged: Error: 2248217, 2248163. See Error Concepts for specifics.
Error message
2248217 Receiver calibration error(s). Receiver DB out of spec. See Long Description 2248217 4b2a diagCoord RrfCalibrator::rcvCal 3779Diagnostic: RRF Calib Iteration: 32 of infinite Action: 1Task ID: 0x9d29e0 Time: 2847174 Object: APS Receiver calibration error(s). Receiver DB out of spec. R1 values that failed: Rx1: 12 10 Rx2: 12 11 10 9 Rx3: 12 11 10 9 Rx4: 12 11 10 Rx5: 12 11 10 9 Rx6: 12 11 10 9 Rx7: 12 11 10 9 Rx8: 12 10 2248163 Receiver calibration validation error. 2004 Error: 2248163 Host: OR3T Process: NSP File: APS:RrfCalibrator::rcvCal Line: 4036 Receiver calibration validation error. Significant measurement discrepancies exist between receivers. First detected: Rx1 vs. Rx2
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2349808 RRF RECEIVER II BOARD See Long Description * A part that does not have a “–2” or “–3” can be replaced with either the same part number or the same number ending with a “–2” or “–3”. If replacing the “–2” or “–3” version, then only the part number with a “–2” or “–3” can be used.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.