Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000738650056260007422

LightSpeed 5.X (80/100kW)

Symptom

Scanner Hardware Stopped Scan with DAS parity errors

Error message

Error: 260007386 DAS Has Detected An Error. Data Parity Error detected in the xx Data Stream(s) on Converter Board xx or the DCB Board. See LD for sample errors Wed Mar 23 15:46:11 2005 Error: 260007386 Host: DAS Process: DCB Firmware File: parity.c Line: 219 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 214/5/1 Type: Helical Exception Level: Sec/Soft Time: 15:46:11:545 Log Series: 660 SCAN ABORT -- DAS Has Detected An Error. Data Parity Error detected in the 1B Data Stream (s) on Converter Board 33 or on Converter Board 41 . NOTE: Before troubleshooting, go through gesys.log (viewlog) and search for all “Data Parity Error” reports in the log. There are 2 items you need to note from the error: 1- Data Parity Error detected in the xx Data Stream and 2- On Converter Board xx or on Converter Board xx Tue Jan 10 09:37:14 2006 Error: 260007386 Host: orp Process: orp_subsys File: parity_16.c Line: 255 Exception Class : Abort Severity : Secondary Function : No System Function Reported Scan Type : Static Scan: 50056/1/1 SCAN ABORT -- DAS Has Detected An Error. Data Parity Error detected in the 3A Data Stream (s) on Converter Board 55 or on Converter Board 71 . Note: Error 260007

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required