Pular para o conteúdo principal
GE Healthcare · CTHiSpeed25380116016176001

Scanner Hardware Stopped Scan WITH View Length Error

Symptom

Scanner Hardware Stopped Scan WITH View Length Error See Long Description for possible error Scanner Hardware Stopped Scan DURING Scan data acquisition WITH Error: 185401 slip-ring view length errors 1 and Error: 253801 View Length Error

Error message

Error: 253801 View Length Error. The size of the views received were incorrect See Long Description for complete error Mon Dec 6 12:12:06 2004 Error: 253801 Suite: RHCT Host: FEP Process: CoordinateMode File: fepin.c 1.61 Method: FEPIN_err_handler Line: 2355 Function: Data Acquisition : Scan Data Acquisition Scan: 16016/2/5 Type: Helical Scan Seq: 176001 FEP DSP 0: View Length Error. The size of the views received were incorrect (HDW status: 0x4006b7) (Data Type: 0x1). View Sequence Number: 6537. Possible causes: Slip Ring buffer terminator boards, SCOM board, or FEP board

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.