Symptom

Multiple recon retries and suspended images during an exam. See long description for details Multiple recon retries and suspended images during an exam. To get the images to recon you would have to resume the queue or reboot the system for the images to recon.

Error message

Error: 230013015 xx DAS views of 100 retrieved from the scan database failed checksums See Long Description for great info on this error Problem: The Following errors all point to a scsi Scan data disk issue. Using the gesys_Ct01.log and iceConsole.log is a must, together both logs point to the Scan data disk chain. Check DIP stats if stats are clean the das data is good. The iceConsole log is a little difficult to read and hard to correlate to specific dates and times. --------------gesys_CT01.log--------------- Sun Jan 13 20:33:00 2002 Error: 245047 Host: gect88 Process: scanmgr File: scan_sequence.cxx Line: 1377 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 30000 Unable to allocate scan files. Sun Jan 13 20:33:05 2002 Error: 230013015 Host: ice Process: rebin File: RebinMDASData.cxx Line: 332 Function: Image Generation : Data Restore Job Type: Prospective Helical Scan: 31/2/1 Image: 31/2/56 Recon Seq Id: 173 xx DAS views of 100 retrieved from the scan database failed checksums <-- DAS Das fails checksum from Scan data disk!!!! Sun Jan 13 20:33:05 2002 Error: 230005024 Host: ice-sm-gate Process: sdcctrl File: sdcctrl_events.cxx Line: 358 imag

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5112299 SCSI cable Replaces 2281056 used in 2.X, 3.X, H3 console (NOT for GOC1 or GOC2 Consoles) See LD for details SCSI cable 5112299 replaces obsolete SCSI cable part number 2281056 used in 2.X, 3.X, H3 console (NOT for GOC1 or GOC2 Consoles) picture below Click here if you need scsi cable for a GOC1 or GOC2 console Part #: 5112299 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PMC SCSI card hard drive cable FRU: Yes Status: Orderable Comments: Added per ECR 2006148, 5112299 is direct replacement for cable 2281056. Please discontinue use of 2281056
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.