GE Healthcare · MRSigna2241545224152322415172241537

Will Not Scan with Gradient Errors

Symptom

Will Not Scan with Gradient Errors.

Error message

2241545 The GIP reported a second Long Overload on the X Axis of the GPM. Tue Apr 20 20:24:34 2004 Error: 2241545 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 653 The GIP reported a second Long Overload on the X Axis of the GPM. Power cycle the main breaker on GPM Power Supply to recover from this condition. If this fault occurs again, swap the GPM axis. Run diagnostics.

2241523 The GIP Reported a Long Overload Fault on the X Axis of the GPM. Tue Apr 20 20:24:33 2004 Error: 2241523 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 2324 The GIP reported a Long Overload Fault on the X Axis of the GPM. Possible causes: X Axis GPM. Try swapping the GPM Axis connections to isolate the problem.->

2241517 The GIP reported a Transistor Fault on the X Axis of the GPM. This Fault Could be Related to Noise. Tue Apr 20 20:24:32 2004 Error: 2241517 Host: lx-mr Process: NSP File: HPC_PORT_ID:GAP/GIP/GP Line: 2218 The GIP reported a Transistor Fault on the X Axis of the GPM. This fault could be related to noise. Possible causes: Cable between the GIP and GPM, X Axis GPM Main Board or Output Shelf.->

2241537 The X Axis of the GPM failed to go to Ready when Commanded. Tu

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
8.3 8.3
Component
2151183 GRADIENT POWER MODULE For SGD. 8651 Is SGD Hi-Slew Gradient Power Module.
Area
MR