SPI To Workspace Interface Module Link Is Not Working

Symptom

SPI To Workspace Interface Module Link Is Not Working System is unable to prescan, APS or MPS, fails with a message of SPI-Workspace Link is not working. The TPS Reset had been done and a workspace shutdown and reboot, but did not help.

Error message

2224666 Auto Prescan failed, SPI-Workspace Interface Module link is not working.

2225385 Workspace Interface Module Initialize failed. See Long Description Mon May 23 08:17:18 2005 Error: 2225385 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_cpb_initialize_fail Line: 4 Workspace Interface Module Initialize failed. If recovery fails, check Workspace Interface Module connections and run Workspace Interface Module diagnostics.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Area
MR

Full cause & step-by-step resolution