Pular para o conteúdo principal
GE Healthcare · MRSigna2223631221800922279722224356

30364 Some or All Images in the Last Scan Did Not Reconstruct

Symptom

Some or All Images in the Last Scan Did Not Reconstruct Some or All Images in the Last Scan Did Not Reconstruct.

Error message

2223631 Internal Software Error. Try to reply to a response packet Wed Oct 27 08:28:18 2004 Error: 2223631 Host: lx-mr Process: ifcc File: mrmailmgr.c Line: 875 Internal Software Error. Try to reply to a response packet.

2218009 Internal Software Error. Unable to send GpcMail packets Wed Oct 27 08:28:18 2004 Error: 2218009 Host: lx-mr Process: ifcc File: IfccScan.cc Line: 7625 Internal Software Error. Unable to send GpcMail packets. Status = -1

2227972 Wait time to receive response from TIR exceeded 20secs...still waiting

2224356 Unable to register scan for install; out of memory Wed Oct 27 08:28:50 2004 Error: 2224356 Host: lx-mr Process: TIR File: tirexec.c Line: 214 Unable to register scan for install; out of memory. Exam/Series/Scan: 8339 7 1

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR