GE Healthcare · Nuclear MedicineInfinia

Service Note 43442_0 May/03 By-passing the Table Mechanical Load Limiter

Symptom

Service Note 43442_0 May/03 By-passing the Table Mechanical Load Limiter There have been incidents where the additional weight of a large patient caused the 18 Kg mechanical Load Limiter to activate prematurely, causing the clutch to activate and unintentional interruption of the acquisition. No Updates

Applies to

Models
Infinia
Product
Infinia I Infinia I Acquisition System Infinia II Infinia II Acquisition System VGH VG and VG Hawkey and Varicam
Component
*Patient Table *Patient Table *Patient Table
Area
Nuclear Medicine