GE Healthcare · CTLightSpeed200107023200109105
LightSpeed 5.X (80/100kW)
Symptom
Scanning aborted with out of memory for scan files. Scanning aborted with out of memory for scan files
Error message
Error: 200107023 A Failure had been detected while calling UIR x_ToScanRequest function or srx_sm_generate_action function See Long Description for error variations Tue Oct 19 09:22:20 2004 Error: 200107023 Host: dgct Process: scanRx File: srx_sma_manage_scan.c Line: 688 Function: Data Acquisition : OC Processing srx_sma_prepare_send_request: A Failure had been detected while calling UIRx_ToScanRequest function. Nov 22 08:55:38 2006 Error: 200107023 Host: ct01 Process: scanRx File: srx_sm_event_notify.c Line: 148 Function: Data Acquisition : OC Processing srx_sm_event_notify: A Failure had been detected while calling srx_sm_generate_action function. Tue Aug 14 16:28:25 2007 Error: 200107023 Host: cccgvct1 Process: scanRx File: srx_sma_manage_scan.c Line: 733 Function: Data Acquisition : OC Processing srx_sma_prepare_send_request: A Failure had been detected while calling SIRx_areScanFilesAvailable function. Error: 200109105 The current prescription requires 290 MBytes of memory for scan files See Long Description for full Error Sun Dec 18 05:20:55 2005 Error: 200109105 Host: 77p_ct_exrm1 Process: scanRx File: SIRx.cxx Line: 5221 Function: Data Acquisition : OC Processing The curre
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 05MW14.5 for all systems with GRE/Xtream consoles on HiSpeed QX/i and LightSpeed 1.X, 2.X, 3.X, 4.X, 5.X Series(QXi, PLUS, ULTRA, LS16, PRO 16, RT). Obsoleted by FMI 25365 deploy FMI 25365 It is the Service person's responsibility to use the appropriate documentation when performing and verifying service. Refer to FMI 25365: 06MW03.5 SW Update for CT-LS 1X/2X/3X/4 DIR 5161945-100, which can be found on the Distributed Technical CDROM/DVD, or online Links to any needed documentation is found in the Product concepts as an attachment called "ProductDocumentLinks"
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.