GE Healthcare · CTLightSpeed26000739111401125991140012600
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors
Error message
Error: 260007391 The Data Acquisition System Has Detected An Error In The -/+12V analog DAS Power Supply See LD for more details Tue Aug 8 07:49:05 2006 Error: 260007391 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 90 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 07:49:05:835 Log Series: 604 The Data Acquisition System Has Detected An Error In The -12V Analog DAS Power Supply.^B Measured: 0 mV , Expected range: -11401 to -12599 mV Tue Aug 8 07:49:05 2006 Error: 260007391 Host: DAS Process: DCB Firmware File: pwrtst.c Line: 90 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 07:49:05:735 Log Series: 604 The Data Acquisition System Has Detected An Error In The +12V Analog DAS Power Supply.^B Measured: 0 mV , Expected range: 11400 to 12600 mV
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5191271 MDAS 16 Fuse Holder Upgrade Kit. Supercedes part number 46-170042P28. See Long Description for more information. Part #: 5191271 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: MDAS 16 Fuse Holder Upgrade Kit FRU: Yes Status: Orderable Comments: Per ECR 2036788, 212008898, 11/30/2006. MDAS Fuse Holder Replacement Kit Contents: Part #: 46-170042P28 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: 30A, 250V PANEL-MOUNT FUSEHOLDER. FRU: Yes Status: Orderable Comments: FLS - FUSE HOLDER PART OF FMI KIT ** FE Assist ** IF FUSE HOLDER P/N 46-170042P28 FAILED, PLEASE ORDER MDAS FUSE FIX ON FAILURE KIT 5191271 (12/12/06 212001261)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.