Infinia II Testing Safety Circuit Operation (Emergency System Check) Recovering From PSD Activation in L-mode…

Infinia II Testing Safety Circuit Operation (Emergency System Check) Recovering From PSD Activation in L-mode

Symptom

Infinia II Testing Safety Circuit Operation (Emergency System Check) Recovering From PSD Activation in L-mode When testing the Safety Circuit the manual instructs you to use the IN key to move the detectors in. Then press a PSD and use the OUT key to move the detectors away. This procedure only works for H-Mode. In L-Mode the detectors move in using the IN key and with a PSD pressed you have to use the Detector UP key to move out. When testing the Safety Circuit the manual instructs you to use the IN key to move the detectors in. Then press a PSD and use the OUT key to move the detectors away. This procedure only works for H-Mode. In L-Mode the detectors move in using the IN key and with a PSD pressed you have to use the Detector UP key to move out. Contributor 212004100 Frastaci December 15 2005

Applies to

Models
Infinia
Product
Infinia II Infinia II Acquisition System
Component
*Gantry Gantry Gantry
Area
Nuclear Medicine

Full cause & step-by-step resolution