Pular para o conteúdo principal
GE Healthcare · CTHiSpeed185110

HiSpeed CT/i 4.X with Indigo and HSA Gantry

Symptom

Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .

Error message

Error: 185110 HW error: Cradle Interference/Limit switches are INVALID Mon Dec 3 12:54:02 2004 Suite: CT01 Host: ETC Proc: Control Error: 185110 File: etcValidIntfMatrix.c 1.8 Method: No Method Line: 58 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 3081 Log Series: 2 HW error: Cradle Interference/Limit switches are INVALID. Interference/Limit switches = 0x4d8. EXTENDED LIMIT: 1 INTF: 0 (0 = NOT active 1 = active)

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-136334p23 -Tilt Limit, Cradle Limit, and Tilt Interference Switches. used on HSA, CTi, 1.x LS Gantries and LS 1X - 4X Zeus table See LD for more info 46-136334p23 -Tilt Limit, Cradle Limit, and Tilt Interference Switches. used on HSA, CTi and 1.x LS Part #: 46-136334P23 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: HOME SWITCH-Omron Z series FRU: Yes Status: Orderable Location in Table:
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.