Pular para o conteúdo principal
GE Healthcare · MRExciteSigna2249149225230122491512252293

Unable to scan with Multicoil Bias Fault & Head & Body T/R Faults, Head and Body Path Open, all at the same time

Symptom

Explanation for Multicoil Bias Fault with Head & Body T/R Faults: Head and Body Path Open Unable to scan using a certain phased array coil. The strange thing is we are always seeing Body Path Open and Head Path Open T/R Faults at the same time as a Multicoil Bias Fault whenever we attempt a scan using this coil. We expect only the Multicoil Bias fault if the problem is with this coil. Why are we also getting the Head & Body T/R Faults? What is the real problem we should be looking at here? See Long Description for the error log. Mon Dec 12 19:42:36 2005 2252293 MRISB1 NSP SCP:DCB 1986 The Driver Module Has Detected a Fault. Details of the Fault can be found in other error messages.-> Mon Dec 12 19:42:36 2005 2252301 MRISB1 NSP SCP:DCB 779 Multicoil Bias Fault Detected - Open Circuit. Try re-seating all coil connectors. MCD1 Channel 2 of 16, counting from 1 Value: 197 mA Limit: 198 mA Mask of all failed channels: 0x2-> Mon Dec 12 19:42:36 2005 2249149 MRISB1 NSP SCP:../TrDriverMdsHWO.cpp 1286 T/R driver fault, head path open Possibly a disconnected cable between the driver board and the head T/R network, or an open diode in the head T/R switch. Mon Dec 12 19:42:36 2005 2249151 MRISB1 NSP SCP:../TrDriverMdsHWO.cpp 1302 T/R driver fault, Body Path open Possibly a disconnected cable between the driver board and the body T/R network, or an open diode in the body T/R switch.

Error message

2249149 T/R driver fault, head path open. 2249149 Host: LDEX Process: NSP File: SCP:../TrDriverMdsHWO.cpp Line: 1286 T/R driver fault, head path open. Possibly a disconnected cable between the driver board and the head T/R network, or an open diode in the head T/R switch.

2252301 Multicoil Bias Fault Detected - Open Circuit. Try re-seating all coil connectors. MCDx Channel x of y, counting from 1 Value x mA Limit y mA See Long Description for Extended Error 2252301 MRI_GB_OW0 NSP SCP:DCB 816 Multicoil Bias Fault Detected - Open Circuit. Try re-seating all coil connectors. MCD1 (or MCD2) Channel 1 of 16, counting from 1 Value: 2 mA Limit: 298 mA Mask of failed channels: 0x1 Ch 1-8 : Legacy MC Conn Ch 17-24 : Conn A Ch 25-32 : Conn B-> Extended Error: This fault was detected by the Driver Module. An open circuit fault occurs when the actual value is less than the pre-defined limit. Check system cables, connectors, coils and configurations. Re-seat all coil connectors. The mask of all failed channels indicates additional channels with the same fault - convert the hex number to a bitmap to decipher the mask, with channel 1 indicated by the least significant bit and channel 16 indicate

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
10.0
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues