PM Head EPIWP prematurely ends and Power Monitor Peak Power Exceeded Faults are visible in Error Log
Symptom
PM Head EPIWP prematurely ends and Power Monitor Peak Power Exceeded Faults are visible in Error Log Effectivity: All Signa 1.5T and 3T EXCITE HD (12.0) MR Systems. Problem Description: After running LVshim the Head PM EPIWP (EPI White Pixel) test will complete quickly and not generate data. Body EPIWP will complete normally and generate data. A review of the error log will show that a Head Power Monitor Peak Power Exceeded Fault ocurred and caused the Head portion of the test to end prematurely. (SMILe Nov. 10, 2005)
Applies to
- Models
- Excite, Signa
- Product
- 3T Excite HD 12.0 software only
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required