Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000650050000

Scanner Hardware Stopped Scan

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260006500 The axial AC Drive reported fault code: F4 - Undervoltage Fault See the LD for the full error code Error: 260006500 The axial AC Drive reported fault code: F4 - Undervoltage Fault Tue Nov 16 13:47:02 2004 Error: 260006500 Host: STC Process: Axial File: AX2fault.c Line: 196 Function: AXIAL CONTROL : axial control Scan: 50000/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 13:47:03:015 Log Series: 117 The axial AC Drive reported fault code: F4 - Undervoltage Fault The Drive internal DC bus voltage dropped below 325V. Possible low voltage condition on 480 VAC in gantry or power interruption. Also, may indicate excessive run or braking power required due to sluggish Motor.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-170021P15 (250V,8A) Fuse used in axial dynamic brake circuit as CT2A1A10F1 and found in FAN ASSY- GANTRY, and F4, F5 on HEMRC Interface Board and other places See long description for details Part #: 46-170021P15 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 008.000A 0250V 3AB SLOW UL/ FRU: Yes Status: Orderable Comments: FLS - FUSE 8 AMP - 250 VOLT 3AB
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.