Pular para o conteúdo principal
GE Healthcare · MRExcite

MDC Ground Current is high with MR System powered off.

Symptom

MDC Ground Current is high with MR System powered off. One of the biggest problems with MR systems has been excessive ground current with the system powered off. Excessive ground current is usually anything over 50 mA. Please remember that this current is on the ground wire from the PDU to the MDC with ALL the breakers OFF in the MDC. It is not uncommon to find this ground current to be as high as 1.5 AMPS.

Applies to

Models
Excite
Product
Profile with Excite
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues