Scanner hardware reset fails
Symptom
Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .
Error message
Error: 178575 Failed to R-HARD the following controllers: ETC. See Long Description for complete error. Error: 178575 Failed to R-HARD the following controllers: ETC. Fri Nov 25 13:10:08 2005 Suite: CHCT Host: SBC0 Proc: scse Error: 178575 File: HWinit.m 1.3 Method: resetControllers Line: 384 Failed to R-HARD the following controllers: ETC. Thu Jun 21 08:45:51 2007 Error: 178575 Host: ct01 Process: fwmgr File: HwSetUp.c Line: 837 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to R-HARD the following controllers: STC ETC OBC. 212002267 06/21/2007 13:15 GMT
Applies to
- Product
- HSA Z Series HSA Z Series HSA Z Series Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-297889G3 Huerikon board..... 46-297889G3 Huerikon board..... 46-297889G3 Huerikon Board Part #: 46-297889G3 Part Type: Not Available Part Weight: 0.00 LBR Repairable: Not Available Returnable: Not Available Hazardous Material: N Description: Not Available FRU: No Status: Non Orderable Comments: FLS - FE: THIS IS HEURIKON BD FOR ZB SITES. GELS - FOR ZB AND ZC RELEASE .UPGRADABLE INTO 46-297889G5 THAT HAVE TO BEUSED AFTER RP1.3 FMI25178 OR FMI25179 ARE INSTALLED.SP11/94 DNA DU TO US CRISIS DO NOT RELEASE WITHOUT OLC AGREEMENT,PCB CPU HEURICON FOR ZB RELEASE
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required