Pular para o conteúdo principal
GE Healthcare · CTHiSpeedLightSpeed182207104947063665430

HiSpeed QX/i - Octane

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 182207 HW error: Alignment error between Encoder and Potentiometer See Long Description for complete error Tue Aug 5 15:05:39 1997 Error: 182207 Host: ETC Proc:Cradle File: cmcCorrErr.c 1.22 3/11/92 Line: 51 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 10494/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:05:39:320 Log Series: 28 HW error: Alignment error between Encoder and Potentiometer. Potentioment Position 70636 (.01 mm) Encoder Position 65430 (.01 mm) Tolerance 2540 (.01 mm)

Applies to

Models
HiSpeed, LightSpeed
Product
HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-297036G2 longitudinal potentiometer See LD for more info 46-297036G2 longitudinal potentiometer ( NOTE the pot/encoder assy is # 46-296234G1 ) Part #: 46-297036G2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: POT ASSY. FRU: Yes Status: Orderable Comments: FLS - PF8 TWICE FOR FE ASSIST COMMENTS...... COLLECTION OF DEFECTIVES START7/23/01 MS5277 SPP CODE CHANGED FROM 2 TO N PER PCN 220431. MSA8429 03/25/02 PF8 AGAIN FOR FA ASSIST COMMENTS. FE ASSIST: IN ORDER TO INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD. ORDER THE "U" PART IF AVAILABLE. IF NO "U" USED PARTS ARE AVAILABLE CARES WILL CHANGE THE ORDER TOA NEW PART. FLS - FE ASSIST (971512) (MS6588) 46-297036G2 IS THE HSA POTENTIOMETER ASSEMBLY WHICH CONSISTS OF A BURNS 5K POT AND A 2.2 MICRO FARAD CAP. THE CAPACITOR IS SOLDER ACROSS THE WIPER AND RETURN TERMINALS. THE 9800 TABLE ELEVATION POT (46-165135P3) IS THE SAME 5K BURNS POT BUT DOES NOT HAVETHE NOISE REDUCING CAPACITOR PLEASE REFER TO SN-T1336 FOR DETAILS REGARDING CAPACITOR AND POT NOISE DO YOU STILL WISH TO ORDER THIS PART?
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues