Pular para o conteúdo principal
GE Healthcare · MRExcite2248356

No Clock Signal Due To an open primary winding on the low-voltage AC line clock transformer

Symptom

Can Not Scan

Error message

2248356 SSI Watchdog Timeout. An SSI Was Not Detected in the Timeout Period. See Long Description Wed Jul 7 09:33:17 2004 Error: 2248356 Host: GEMRI_1Process: NSP File: SCP:sct_wdog_exp Line: 852 SSI Watchdog Timeout. An SSI was not detected in the timeout period. The PSD or AGP might not be running. It may be necessary to perform a TPS reset. If TPS Reset doesn't help then AC 60Hz clock may be missing. Extended Error: Check the 60Hz clock signal from PDM to STIF Board in CAM chassis. Run CAM Data Acquisition SRF Diagnostics.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues