GE Healthcare · CTLightSpeed2600065005003826000683650000
LightSpeed 4.X - GOC2
Symptom
Gantry will not rotate or scan, alignment lights will not turn ON
Error message
Error: 260006500 Axial AC Drive reported fault code: F60 - Encoder loss. Encoder Loss indicates that the axial drive can not sense the axial encoder See long description for full error Sun Jan 08 11:00:27 2006 Error: 260006500 Host: STC Process: Axial File: AX2fault.c Line: 196 Function: AXIAL CONTROL : axial control Scan: 50038/8/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 11:00:26:535 Log Series: 35 The axial AC Drive reported fault code: F60 - Encoder Loss This indicates that the axial drive can not sense the axial encoder. Possible causes are: Encoder, cabling, unplugged axial control board or improper axial drive encoder jumper setting. Error: 260006836 The Encoder and the Drive MUST be Sync'd/Referenced. See the LD for the complete error code Tue Nov 16 14:20:36 2004 Error: 260006836 Host: STC Process: Axial File: amc_exec.c Line: 2132 Function: AXIAL CONTROL : axial control Scan: 50000/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 14:20:36:955 Log Series: 69 Axial Motor Control (AMC) Logic will NOT Comand the DRIVE until the Gantry motion is STOPPED. The Encoder and the Drive MUST be Sync'd/Referenced. (Cur State: 10 Cmd: 0x170a0048) HW/SW Er
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2218570 BOX OF FUSES for HSA, CTI, and LS 1X - 4X Systems See long description for details 2218570 BOX OF FUSES for HSA, CTI, and LS 1x - 4x Systems Fuse KIT
- Area
- CT
Full cause & step-by-step resolution