GE Healthcare · CTHiSpeed214052365316392118328950001
Scanning aborted with unexpected Loss Of High Voltage.
Symptom
Scanning aborted with unexpected Loss Of High Voltage.
Error message
Error: 214052 The HV status indicator has reported a premature termination of X-Rays See Long Description for complete error code Tue Oct 10 21:27:05 1995 Suite: CT01 Host: OBC Proc: Diag Exec Error: 214052 File: isr3_handler.c 1.28 Method: obc_isr3_handler Line: 187 Function: Diagnostics XRAY GENERATION : Rotating HVON Control Test Name: Manual KV Unique ID: 36531 Iteration: 1 Exc. Level: Hard Time: 63921 Error Type: Hardware The HV status indicator has reported a premature termination of X-Rays. Technic: 120KV 0mA Exposure duration: 83mS Expected: 10000mS KV address: FFCFF9H HV indicator: Bit D0 Value 0 Exposure Command: Bit D2 Value 1 Error: 183289 Unexpected Loss Of High Voltage Wed Aug 4 20:21:15 2004 Error: 183289 Host: OBC Process: Generator File: gen xray.c 1.31 Line: 528 Function: XRAY GENERATION : Rotating HVON Control Scan: 50001/1/1 Type: Axial Exception Level: Pri/Most Severe Time: 20:21:16:325 Log Series: 59 Unexpected Loss Of High Voltage. .
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-170021P92 SEMICONDUCTOR FUSE - 60A 700V Part #: 46-170021P92 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: SEMICONDUCTOR FUSE FRU: Yes Status: Orderable Comments: FLS - CLASS 1 FRU INACTIVE TO MFG. CJB 1/94 60A 0300V LIMTD RANGE SLO W UL/C DUAL ELEMENT. (REVIEWED TK, 7/91 - OK FUS RAP 60A 700V FBP60 FLS - FE ASSIST: IN ORDER TO INCREAS INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCRESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD. ORDER THE U PART WHEN AVAILABLE. IF NO U USED PARTS ARE AVAILABLE CARES WILL CHANGE THE ORDER TOA NEW PART. SEMICONDUCTOR FUSE FLS - CLASS 1 FRU INACTIVE TO MFG. CJB 1/94 60A 0300V LIMTD RANGE SLO W UL/C DUAL ELEMENT. (REVIEWED TK, 7/91 - OK FUS RAP 60A 700V FBP60
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.