GE Healthcare · CTLightSpeed18288416417182885
Operator reported that table foot pedals had stopped working
Symptom
Table foot pedals malfunction. Table foot pedal malfunction.
Error message
Error: 182884 The Load Foot Pedal Is STUCK. See LD for full sample error message. Mon Oct 17 20:14:50 2005 Error: 182884 Host: ETC Process: Gantry Display File: gd_tnc_erlog.c Line: 623 Function: No System Function Reported Scan: 16417/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 20:14:50:925 Log Series: 240 The Load Foot Pedal Is STUCK. Error: 182885 The Home Foot Pedal Is STUCK. See LD for sample error message. Mon Oct 17 20:14:51 2005 Error: 182885 Host: ETC Process: Gantry Display File: gd_tnc_erlog.c Line: 631 Function: No System Function Reported Scan: 16417/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 20:14:51:975 Log Series: 241 The Home Foot Pedal Is STUCK.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 405L.5_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console (2378862) 405L.5_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console (2378862) 405L.5_H4.2M4 for LightSpeed 4.X (LightSpeed 16) with Console GOC2 (w/Linux PC) (2315653) or Xtream (GRE) Console (2378862) Supports GOC2/ GRE LightSpeed 16 (4.X) Obsoleted by FMI 25338
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.