Pular para o conteúdo principal
GE Healthcare · CT182912182904

Lightspeed 1.X - Octane

Symptom

Gantry controls would not work when software was rebooted See the long description for the errors Thu Sep 16 06:48:08 2004 Error: 182904 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 370 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Unknown Exception Level: Pri/Most Severe Time: 06:48:00:070 Log Series: 20 A table side key is either stuck or pressed during initialization or just after Emergency Off released. Port value 0x0410. Disabling those keys detected being stuck Thu Sep 16 06:48:08 2004 Error: 182906 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 370 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 06:48:00:070 Log Series: 20 Gantry/Tableside Up Elevation button is non-functional. Use the Foot pedals. Thu Sep 16 06:48:08 2004 Error: 182912 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 406 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 06:48:00:075 Log Series: 20 Gantry/Tableside FAST button is non-functional

Error message

Error: 182912 Gantry/Tableside FAST button is non-functional See the LD for the full error code Error: 182912 Gantry/Tableside FAST button is non-functional Thu Sep 16 06:48:08 2004 Error: 182912 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 406 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 06:48:00:075 Log Series: 20 Gantry/Tableside FAST button is non-functional

Error: 182904 A table side key is either stuck or pressed during initialization or just after Emergency Off released. See the Long Description for the full error code Thu Sep 16 06:48:08 2004 Error: 182904 Host: ETC Process: Table Controls File: tc_exec.c 1.27 1/15/96 Line: 345 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 06:48:00:070 Log Series: 20 A table side key is either stuck or pressed during initialization or just after Emergency Off released. Port value 0x0410. Disabling those keys detected being stuck

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues