Pular para o conteúdo principal
GE Healthcare · CTHiSpeed1832621831825000144858

HiSpeed QX/i - Octane

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 183262 Backup contactor opened without firmware command See the long description for the full error Mon May 17 12:20:18 1999 Error: 183262 Suite: CT01 Host: OBC Process: Generator File: ctvrc_manager.c 1.8 Method: No Method Line:823 Function: XRAY GENERATION : 550 VDC Backup Contactor Intlk Scan: 7073/3/15 Type: Axial Exception Level: Pri/Most Severe Time: 12:20:18:495 Log Series:513 Backup contactor opened without firmware command.

Error: 183182 I/O board pilot relays are closed. PDU pilot relay not energized. HVDC Rail Voltage is xxxx See Long Description for more details. Error: 183182 I/O Bd. pilot relays are closed. PDU pilot relay not energized. HVDC is xxxx. ** Note the HVDC Rail Voltage. It is important in troubleshooting Tue Sep 21 06:13:59 2004 Error: 183182 Host: OBC Process: Generator Function: SYSTEM POWER CONTROL: HVDC Control Scan: 50001/1/1 Type: Axial Exception Level: Pri/Most Severe I/O board pilot relays are closed. PDU pilot relay not energized BCSTAT = 0 HVDC Rail voltage = 0V Sun Dec 12 18:11:57 2004 Error: 183182 Host: OBC Process: Generator File: gen_rotor.c 1.39 Line: 488 Function: SYSTEM POWER CONTROL : HVDC Control Scan: 44858/1/1 Type: Scout Exce

Applies to

Models
HiSpeed
Product
HiSpeed QX/i - Octane Always MDAS4 101I.2_HSQXIM5 101.4_HSQXIM4 HiSpeed QX/i - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 101I.2_HSQXIM5 101.4_HSQXIM4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.