GE Healthcare · CTLightSpeed182825
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
Gantry Display blank. Gantry Display blank
Error message
Error: 182825 HW/SW error: Communication Problems with the GANTRY DISPLAY. Open Long Description for full error Wed Dec 22 20:49:26 2004 Error: 182825 Host: ETC Process: Gantry Display File: gd_tnc_driver.c Line: 471 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 20:48:40:485 Log Series: 30 HW/SW error: Communication Problems with the GANTRY DISPLAY. Number of Communication Timeout errors: 0. Number of NACK's errors: 3. Number of DUART errors: 0. Numver of Checksum errors: 0.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2381519 GANTRY DISPLAY/CONTROL CABLE TO PC 2381519 GANTRY DISPLAY/CONTROL CABLE TO PC Flash Cable Pinout PART # 2381519 Should your flash cable labels wear off, the following chart provides the pinout. Note: The pins that are not shown in Table 1 are unused. Table 1: Flash Cable Pinout CONTROL PANEL END PC END SIGNAL Pin 1 Pin 5 GND Pin 3 Pin 3 Tx out of PC Pin 4 Pin 2 Rx into PC Cable style: 9 pin female to 9 pin female, serial style PC cable. Cable length: 10.5 feet (3.2 meters).
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.