GE Healthcare · CTHiSpeed1100311002
27706 Certain exams fail to network to PACS.
Symptom
Certain exams fail to network to PACS. Exams will just pause in network queue.
Error message
Error: 11003 Failure to get DCM event: lost transport connection (114) Fri Jul 23 08:28:45 2004 Error: 11003 Host: lightspeed04 Process: dcserver File: dcspush.c Line: 2514 Failure to get D From viewlog: Fri Jul 23 08:28:45 2004 Error: 11003 Host: lightspeed04 Process: dcserver File: dcspush.c Line: 2514 Failure to get DCM event: lost transport connection (114) Error: 11002 DCM low level error: Network error (508): lost transport connection (114) Fri Jul 23 08:28:45 2004 Error: 11002 Host: lightspeed04 Process: dcserver File: dcscommon.c Line: 227 DCM low level error: Network error (508): lost transport connection (114) From viewlog: Fri Jul 23 08:28:45 2004 Error: 11002 Host: lightspeed04 Process: dcserver File: dcscommon.c Line: 227 DCM low level error: Network error (508): lost transport connection (114)
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 6.3.1 for HiSpeed CT/i with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.3.1 6.3.1 for HiSpeed CT/i with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.3.1 6.3.1 for HiSpeed CT/i with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.3.1 Class C Class M Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2276664-3 6.3.1M4 HiSpeed CT/I Class C Software 2276665-3 6.3.1M4 HiSpeed CT/I Class M Software 2276666-3 2276663-3 6.3.1M4 HiSpeed CT/I App SW 6.3.1 Rhapsode Product 6.3.1 10/29/01 Only for Systems with Remote Tilt Option
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.