Pular para o conteúdo principal
GE Healthcare · CT18751818751750510

CT/e Single Irix or Linux

Symptom

Image Artifacts with Uncorrected View Number of BADRAW ; 1, ImageId found in the error log - document in RFS - critical issue Image Artifacts with Uncorrected View Number of BADRAW ; 1, ImageId found in the error log

Error message

Error: 187518 Uncorrected View Number of BADRAW 187518 Uncorrected View Number of BADRAW; 1, ImageId Wed Nov 24 15:26:17 2004 Error: 187518 Host: ahnkct Process: ReconJob File: nreIn.c Line: 688 Uncorrected View Number of BADRAW ; 1, ImageId ; 1950/4/74 If the error is UNCORRECTED, then FEC noticed the error but was UNABLE to correct it - this will result in a blown image artifact If the DASINF is able to correct the data with Forward Error Correction then there will not be any artifacts and the error will be: Fri Mar 4 11:35:28 2005 Error: 187517 Host: ahnkct Process: ReconJob File: nreIn.c Line: 692 Corrected View Number of BADRAW ; 1, ImageId ; 2466/2/1 When running Das Data Transfer Test the error will be: Wed Mar 9 17:13:25 2005;DAS xfer;Raw A#12(ScanTime 1.0)/25 (Loop #6/25) Compare Error ** Raw:50510. 6.12, View.:990, Ch.:101 (3/137) Wed Mar 9 17:16:34 2005;DAS xfer;Raw B#16(ScanTime 1.0)/25 (Loop #8/25) Compare Error ** Raw:50510. 8.16, View.:1194, Ch.:242 (4/191)

Error: 187517 Corrected View Number of BADRAW See LD for complete error 187517 Corrected View Number of BADRAW; 1, ImageId Wed Nov 24 09:23:23 2004 Error: 187517 Host: ahnkct Process: ReconJob File: nreIn.c Line:

Applies to

Product
CT/e, CT Prospeed AI/FI Irix or Linux See long description for sale information Product Name: CT/e Single CT Scanner: Irix Platform A list of types. (a Product/hardware table) N/A An On-site / Remote identification aid . (Example: check_config) N/A A Table of options (included, optional, new introductions, …) N/A A list of Software that has run on this product (Product / Software Compatibility Table) 3.01 / 3.02 / 3.03 Linux CTe Product photographs N/A Marketing Information N/A Documents: Always use the most appropriate documentation for your system. Documentation can be found on CT Service CDROM/DVD, or online at FCT Service Engineering Tech Pubs Website, Common Document Library, and My Workshop. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found in the attachment Always use the most appropriate documentation for your system.
Component
2245896-2 HSDCD RX WITH ESD PROTECTION THIS PN REPLACES 2245896 If you have the old style ETC slip ring this part IS NOT FOR YOU. ETC uses 2167012 rf receiver See long description for more info Part #: 2245896-2 Part Type: GP CONSUMABLES Part Weight: 0.17 LBR Repairable: No Returnable: No Hazardous Material: N Description: HSDCD RX WITH ESD PROTECTION FRU: Yes Status: Orderable Comments: FLS - THIS PN REPLACES 2245896. BEFORE USING THE -2 VERSION, USE UP EXISTING STOCK OF 2245896. ADDED PER PCN 220414, BY MS20544, CSV ON 1/17/02. VC 611168 = SCHLEIFRING
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.