GE Healthcare · CTHiSpeed187517188415
System intermittently failed to scan, which usually occurred at the first several scanned patients after customer power on system in the morning
Symptom
System intermittently failed to scan, which usually occurred at the first several scanned patients after customer power on system in the morning System intermittently failed to scan , which usually occurred at the first several scanned patients after customer power on system in the morning
Error message
Error: 187517 Corrected View Number of BADRAW See LD for complete error 187517 Corrected View Number of BADRAW; 1, ImageId Wed Nov 24 09:23:23 2004 Error: 187517 Host: ahnkct Process: ReconJob File: nreIn.c Line: 692 Corrected View Number of BADRAW ; 1, ImageId ; 1942/2/20 If the error is CORRECTED, then FEC noticed the error and was able to completely restore the data. Customer will notice nothing. Error: 188415 Continuous Bad Raw detected and Scan is going to aborted. Check status LED on DASIFN See Long Description for Full Error. Tue Aug 16 17:28:41 2005 Error: 188415 IRIS RawStore NdbAccess.c 299 DASIFN Error: Continuous Bad Raw detected and Scan is going to aborted. Check status LED on DASIFN
Applies to
- Models
- HiSpeed
- Product
- HiSpeed YMS Series HiSpeed Series DXi, FXi, LXi, ZXI, and NXI Including Single / Twin detector and Glass / Metal Tubes. See Long Description for Details Subsystem/Components DXI (NP) FXI,LXI (NP+) ZXI (NP++ NXI (NP+ Twin NXI (NP++ Twin OC O2 running Unix or upgraded to Linux PC Gantry Alighnment Lights Halogen Laser Laser Laser Laser Mechanics Common Electronics firmware is different twin Det/Das Twin Det/DAS Table IMS no Yes Yes Yes Yes Generator Jedi Jedi NP++ Jedi Jedi NP++ Jedi Tube D3142T D3142T D3152T D3142T D3152T PDU Common Options List for V/R 5.5, 6, 7, 9, and 10 System 3D Power 440ma Max Image Disk 20000 Smart Prep Fast Scam 0.7 sec Performix 530 cooling Smart Recon Fast Scan 0.8 sec Asymmetric Scan SmartView Fast Scan 1.0 sec Cardiac gating DemtaScan Helical 30 Max Smart Addition Navigator Helical 60 Max Respiratory Gating Helical Plus Helical 120 Max Reformat Cpmmect {rp Raw Data 1000 max VolumeRendering2 ImageDosk Remote Tilt CT Colono Pioneer Mod Enhanced Cooling The following Options name changes Option (V5.5x,V6.0x,V7.x,V9.x,V10x) Option (V8.1x and V11.x) 3D >> Volume Analysis2 Navigator >> Navigator2,VolumeRendering2,VolumeAnalysis2 CtPerfusion >> CTPersusion2 Reformat >> Functool Options V8.1x and V11.x VolumeAnalysis2 Power 440ma ImageDisk 2000 Helical Plus Smart Prep Fast Scan 0.7 sec Performix 530 cooling Connect Pro Smart Recon Fast Scan 0.8 sec Asymmetric Scan ImageDisk SmartView Fast Scan 1.0 sec Cardiac Gating Pioneer MOD* DentaScan Helical 30 Smart Addition Power 200 ma max Navigator2/VolumeRendering2/ VolumeAnalysis Helical 60 Respiratory Gating Power 250 ma Max Power 300 ma Max Helical 120 Thickness 0.5mm CTColon Power 350 ma Max Raw data 1000 CtPerfusion2 AdvVesselAnalysis Remote Tilt Extended FOV Functool Enhance Cooling Performix Tube VolumeRendering2
- Component
- 2241209–2 : DASIFN2 used on Hispeed NX/I Linux See Long Description Part #: 2241209-2 Part Type: GP CONSUMABLES Part Weight: 0.30 KGM Repairable: No Returnable: No Hazardous Material: N Description: P2011LN DASIFN2 W/PANEL ASSY FRU: Yes Status: Orderable Comments: FLS - YMS EBISU+ CARDIAC GATING OPTION. P2011LN DASIFN2 W/PANEMS26561 DSM 3/11/01
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required