GE Healthcare · CTLightSpeed18241425267
LightSpeed 3.X (Ultra, Plus, QXi) - Octane
Symptom
Table elevation stopped going up with Current limit fault detected by hardware.
Error message
Error: 182414, Current limit fault detected by hardware. See LD for complete error Error: 182414, Current limit fault detected by hardware Wed Feb 23 21:14:53 2005 Error: 182414 Host: ETC Process: Elevation File: emcCurLimit.c 1.20 11/13/92 Line: 70 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 25267/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 21:14:53:830 Log Series: 445 HW error: Current limit fault detected by hardware. Current Position 4946 (.1 mm) Elev Direction: 0 (0 = Up 1 = Down).
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-288170G1 ELEV/TILT AMP See LD for more 46-288170G1 ELEV/TILT AMP Part #: 46-288170G1 Part Type: GP CONSUMABLES Part Weight: 1.36 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEV/TILT AMP FRU: Yes Status: Orderable Comments: FLS - B: THE PROBLEM IS USUALLY WITHTHE 24 V RELAY. PLEASE HAVE THE FE TRY PART 2120791-9. MSA6521 5/29/01 DELISTED T3202CF 12/00 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA FE ASSIST: IN ORDER TO INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.